Philips Respironics V60 Air Inlet Filter Replacement

ID: 132280

Description: The air inlet is located on the right side...

Steps:

  1. Press and release the ON/Shutdown key, opening the Shutdown window in the screen.
  2. Select Ventilator Shutdown, which will shut down the ventilator.
  3. Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.
  4. Step image Step image
  5. Disconnect the ventilator from the AC power.
  6. Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.
  7. Step image
  8. Remove the inlet filter by pinching it out of the recess in the bracket.
  9. Install a new air filter by tucking it into the recessed area.
  10. Step image