Newport HT70 Air Inlet Filter Replacement
ID: 132298
Description: According to the service manual the Air Intake...
Steps:
- The momentary-type power switch is located on the left side rear of the ventilator along the bottom edge. Press the power switch once.
- A message appears in the screen that prompts you to Press Accept to Shutdown or Press Cancel to Ignore.
- Pressing the Accept button will power off the ventilator.
- Pressing Cancel will return the ventilator to its previous state.
- Press Alarm Silence to silence the Shut Down alarm.
- Disconnect the ventilator from the AC power.
- The inlet air filter is located on the right-hand side of the device.
- Turn the knurled screws counter-clockwise to remove.
- Slide the inlet filter cover assembly toward the back of the device and lift it out.
- Remove the Air Inlet Filter Gasket by carefully pulling it away from the side of the device.
- Replace the air inlet filter.
- The service manual makes the following note: To reattach Air Inlet Filter Gasket, use adhesive NMI part number ADH110P. Apply the adhesive to the rear surface of the gasket avoiding the slotted areas. Align the tab in the gasket with the slot on the device and press firmly and allow curing overnight before returning to service.
- Slide the air filter cover back on.
- Turn the knurled screws clockwise to insert.