Tektronix TBS1062 Oscilloscope Disassembly
ID: 155060
Description: There is a manual for this oscilloscope:...
Steps:
- Remove all eight plastic buttons.
- Remove the five Torx T15 screws.
- Remove the front panel by pulling it off
- Remove the power plug from the cover
- Firmly grasp the ON/OFF switch and pull it straight out of the box.
- Press the button into the ON/OFF button recess until it clicks into place.
- Remove the two black plastic plugs from the bottom.
- Remove the five Torx T15 screws.
- Remove the back cover.
- Even when disconnected, the power supply can retain a dangerously high voltage. Avoid contact with the rear of the board or other components. Use insulated electrician's gloves. Ensure that the capacitors are discharged before touching them.
- Disconnect the J101 power supply connector from the motherboard.
- Disconnect the earth connection.
- Remove the 5x Torx T15 screws.
- Remove the power supply board.
- Disconnect the black cable from the motherboard
- Remove the 5x Torx T15 screws
- Pull the control board plate
- Loosen the three nuts holding the signal connectors.
- Loosen the three T15 Torx screws.
- Remove the cables.
- Remove the main board.
- Disconnect the ribbon cable and the LCD backlight cable from the motherboard.
- Loosen the 4x 8mm nuts from the display supports.
- Remove the LCD display.