Tektronix TBS1062 Oscilloscope Disassembly

ID: 155060

Description: There is a manual for this oscilloscope:...

Steps:

  1. Remove all eight plastic buttons.
  2. Remove the five Torx T15 screws.
  3. Remove the front panel by pulling it off
  4. Step image Step image Step image
  5. Remove the power plug from the cover
  6. Firmly grasp the ON/OFF switch and pull it straight out of the box.
  7. Press the button into the ON/OFF button recess until it clicks into place.
  8. Remove the two black plastic plugs from the bottom.
  9. Remove the five Torx T15 screws.
  10. Remove the back cover.
  11. Step image Step image
  12. Even when disconnected, the power supply can retain a dangerously high voltage. Avoid contact with the rear of the board or other components. Use insulated electrician's gloves. Ensure that the capacitors are discharged before touching them.
  13. Disconnect the J101 power supply connector from the motherboard.
  14. Disconnect the earth connection.
  15. Remove the 5x Torx T15 screws.
  16. Remove the power supply board.
  17. Step image Step image Step image
  18. Disconnect the black cable from the motherboard
  19. Remove the 5x Torx T15 screws
  20. Pull the control board plate
  21. Step image Step image
  22. Loosen the three nuts holding the signal connectors.
  23. Loosen the three T15 Torx screws.
  24. Remove the cables.
  25. Remove the main board.
  26. Step image Step image Step image
  27. Disconnect the ribbon cable and the LCD backlight cable from the motherboard.
  28. Loosen the 4x 8mm nuts from the display supports.
  29. Remove the LCD display.
  30. Step image Step image Step image