How to Restore Suction on a Spectra S1 Breast Pump
ID: 206031
Description: Loss of suction is one of the most common...
Steps:
- Wash hands with soap and warm water prior to touching the pump parts.
- Disassemble removable pump parts (flanges, tubing, valves, back flow protectors).
- Inspect tubing for condensation, milk residue, holes, or loose connections.
- Run tubing through fingers to identify deformities.
- Use a backlight to look for moisture, mold, and/or discoloration.
- Check duckbill valves for stretching, tears, or warping; replace if worn.
- One way to effectively check the seal on a duckbill valve is to fill it with water and see if water leaks out.
- Inspect membrane/backflow protectors for condensation, cracks, poor seal, or incorrect placement.
- Reassemble parts carefully, ensuring tight seals and correct alignment.
- Reattach duck bills to the bottoms of each flange. Reassemble backflow membranes. Attach membranes to port on the back of each flange.
- Using one of the tubing pieces, attach one end to your backflow membrane and the other end to the pump.
- Test suction using one side at a time to isolate potential issues.
- Use the built-in cover over the port you aren’t testing and test each side of parts one at a time to identify unilateral problems.
- Replace additional components (tubing or protectors) if suction is still weak.
- Yellowing of components over time can be normal and does not necessarily indicate replacement is necessary. However, older parts are more susceptible to warping/cracking which may warrant replacement.