Cen-Tech Seven Function Digital Multimeter Ball Bearings Replacement

ID: 32071

Description: If the dial cannot spin then it is likely that...

Steps:

  1. Place the meter face down, and the back panel and screws can be seen.
  2. Step image
  3. Unscrew the two 9mm screws on the back panel using a #0 Phillips screwdriver.
  4. Step image Step image
  5. Remove the back panel of the device by lifting upward.
  6. Remove the battery from the meter before continuing.
  7. Step image
  8. Remove the four 7mm Phillips screws holding the circuit board in place.
  9. Step image Step image
  10. Gently lift the circuit board away from the device casing.
  11. Remove the battery after removing the back cover to reduce the chances of short-circuiting the circuit board.
  12. Step image
  13. The dial is located in the central part of the device.
  14. Step image
  15. Carefully remove dial from device by lifting upward.
  16. Clean existing dial or replace if necessary.
  17. Take care not to lose the small ball bearings.
  18. Step image
  19. Locate ball bearings. Carefully remove.
  20. If both ball bearings are still present, try lubricating ball bearings with multi-use lubricant.
  21. If one or both ball bearings are missing, simply replace.
  22. Step image